Author: Release date: 2017-03-29 09:29:27 Source: Shanghai Cixi View: 123
Although the name of turbine flowmeter is similar, vortex flowmeter is different from turbine flowmeter to a great extent. For example, we will introduce the requirements and possible problems of vortex flowmeter for straight pipe section. Now we will introduce it to you.

The first problem is to look at the requirements of vortex flowmeter for straight pipe section

1. The vortex flowmeter has certain requirements for the upstream and downstream straight pipe sections on the installation point, otherwise the measurement accuracy will be affected.

2. If there are piston or plunger pumps, piston or roots fans and compressors in the upstream of vortex flowmeter, there should be no less than 25d equal diameter straight pipe section in the upstream of flowmeter, and no less than 5D equal diameter straight pipe section in the downstream.

3. If there is a reducing pipe upstream of the installation point of the vortex flowmeter, the upstream of the flowmeter should have equal diameter straight pipe section of no less than 15d, and the downstream should have no less than 5D equal diameter straight pipe section.

4. If there is a 90 ° elbow upstream of the installation point of vortex flowmeter on the same plane, there should be no less than 25d equal diameter straight pipe section upstream of the flowmeter, and no less than 5D equal diameter straight pipe section downstream of the flowmeter.

5. If there is a 90 ° elbow or downcomer upstream of the installation point of the vortex flowmeter, the upstream of the flowmeter should have equal diameter straight pipe section of no less than 20d, and the downstream should be no less than 5D equal diameter straight pipe section.

6. The flow regulating valve or pressure regulating valve shall be installed at a distance of 5D downstream of the flowmeter as far as possible. If it must be installed at the upstream of the flowmeter, there shall be no less than 25d equal diameter straight pipe section in the upstream of the flowmeter, and no less than 5D equal diameter straight pipe section in the downstream.

7. If there is a diffuser upstream of the installation point of the vortex flowmeter, the upstream of the flowmeter should have an equal diameter straight pipe section no less than 18D (D is the pipe diameter), and the downstream should be no less than 5D.

Second, let's look at the possible problems of vortex flowmeter

1. Problem: since the temperature and humidity of coke oven gas are high when leaving the factory, there will be moisture in the gas transmission process. The gas flow drives the water to fluctuate back and forth, thus forming a pulsating flow. When the vortex flow sensor is in this fluid state, the output data is large or small, which can not reflect the production status.

Solution: regularly drain the pipeline, especially the water in front of the straight pipe section. According to the specific situation, special personnel shall be set to discharge regularly, so as to reduce the moisture in the metering pipe section as much as possible, and eliminate the pulsation in the fluid to the maximum extent.

2. Problem: vortex flowmeter is not suitable to be installed in strong vibration situation, which is well known by users. However, when magnetic field changes frequently, vortex flow sensor will measure signal output higher than normal value. Practice has proved that in the field without gas flow, when the vortex flow sensor is in the changing magnetic field, the vortex flowmeter sensor will sense an error signal and output at the moment of magnetic field change. When the change is over and the instrument is in a stable magnetic field, the instrument will output a normal signal.

Solution: in the process of instrument installation and connection, it is necessary to ensure the accuracy of each link, including the site inspection before installation, instrument wiring during installation, system grounding wire, etc., so as to ensure the detection of real data and accurate output.

3. Problem: the pressure line is not solid in the process of instrument wiring, resulting in intermittent signal transmission.

Solution: for the measurement system in operation, the method of "comparative confirmation" and "alternative method" can be used to confirm and eliminate the faults of metering instruments in operation.

4. Problem: the coke oven gas has many impurities, easy to crystallize, and the impurities condense on the sensor head, resulting in inaccurate measurement. With the increase of temperature, the impurities volatilize, the sensitivity increases and the signal increases; on the contrary, it decreases. As a result, the data is unstable.

Solution: clean the instrument as a whole regularly. If necessary, purge the sensing head of the instrument to avoid the condensation of impurities at the sensing head. In the cold season, adding heat tracing device in the metering straight pipe section and instrument part is also beneficial to alleviate the condensation of impurities at the metering instrument.

Whether the vortex flowmeter is selected or installed and used, it is very worthy of the attention of the majority of users. If these problems are not done well, it may face difficulties in use.

The above is the Shanghai Cixi vortex flowmeter on the straight pipe section of the introduction, hope to help you, if you still related product problems, you can call the page of the consultation! The company provides 24-hour service!